1.
Journal of Tropical Medicine ; (12): 45-48,25, 2005.
Article in Chinese | WPRIM | ID: wpr-579367

ABSTRACT

Objective To evaluate the identification and counting effeciency of nucleated red blood cells by Sysmex XE-2100 Hematology Analyzer. Methods Accurancy: nucleated red blood cells were counted from 38 specimens by Sysmex XE-2100 Hematology Analyzer and compared with the number counted under microscopy. Precision: 3 specimens with different values were counted for the nucleated red blood cells 10 times by Sysmex XE-2100 Hematology Analyzer and compared with the number counted under microscopy. The CV(% ) value was estimated.Results There was insignificant difference between the results obtained from Sysmex XE-2100 Hematology Analyzer and those under microscopy. In the T-test, P >0.05, r=0.9893(P< 0.01).CV(% ) were 8.1% and 15.8% . It means that the Sysmex XE-2100 is more precise in analyzing the nucleated red blood cells than that under microscopy. Conclusion The nucleated red blood cells count by Sysmex XE-2100 is accurate and fast to obtain clinical data.

2.
Journal of Tropical Medicine ; (12): 786-789,799, 2005.
Article in Chinese | WPRIM | ID: wpr-571786

ABSTRACT

Objective To evaluate the influence of the storing time of venous blood samples on the differential count of WBC by Sysmex XE-2100 hematology analyzer. Methods At room temperature, the precision of the differential count of WBC by Sysmex XE-2100 hematology analyzer were tested. 38 samples were taken the differential count of WBC by Sysmex XE-2100 hematology analyzer after stored for 0, 2, 4, 8, 24 and 48 h. Differential count of WBC was also taken under microscope for comparison. Results The precision of differential count of WBC by Sysmex XE-2100for all the samples was in the allowable range. The correlation coefficient of the differential count of WBC by two methods for neutrophils, lymphocytes, monocytes, eosinophils and basophils were 0.9859, 0.9775, 0.8053, 0.8695and 0.5243 (P<0.01). There was insignificant difference in the test at 8h, very significant difference of MONO and EOS at 48 h. Differences of EOS, MONO between two methods were significant increased at 8 h. Conclusion At room temperature, the differential count of WBC of venous blood samples by Sysmex XE-2100 hematology analyzer should finish within 8 h.

SELECTION OF CITATIONS
SEARCH DETAIL